Instruction manual
C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-27
tape counter
This command reports the number counts which corresponds to the filter
tape movement. The following example reports that the tape count is 250.
Send: tape counter
Receive: tape counter 250
vac pres
This command reports the current vacuum pressure for the detector, in
mmHg. The following example reports that the vacuum pressure is 100
mmHg.
Send: vac pres
Receive: vac pres 100.0 mmHg
clr records
This command will clear all lrecs and srecs that have been saved.
Send: clr records
Receive: clr records ok
set clr lrecs
set clr srecs
These commands will clear only the lrecs or only the srecs that have been
saved. The following example clears srecs.
Send: set clr srecs
Receive: set clr srecs ok
set copy sp to lrec
set copy sp to srec
set copy sp to stream
These commands copy the current selections in scratch pad (sp) into the
lrec, srec, or streaming data list.
The scratch pad is a temporary memory area which is used to set up lists of
selections for lrec, srec, or streaming data items. The user can copy any of
these lists to the scratch pad, modify individual elements in the list, then
save the scratch pad back to the original list. For more information on how
to edit the scratch pad, see the “sp field” command.
The following example copies the current list in scratch pad into the lrecs
list.
Send: set copy sp to lrec
Receive: set copy sp to lrec ok
Datalogging