C-Link Protocol Commands
Diagnostics
B-26 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Send: io bd
Receive: io bd yes
mass
This command reports the presence of mass. The following example
reports that the mass is 0.1 mg/m
3
.
Send: mass
Receive: mass 0.100E+00
neph bkg
This command reports the current nephelometer background values stored
on the nephelometer interface board. The following example reports the
nephelometer auto-ranging background values.
Send: neph bkg
Receive: neph bkg1 0.00272 0.00272 0.01099 0.04351 0.17430
0.65535
neph mode
This command reports the current operating mode of the nephelometer.
The following example shows that the nephelometer is in sample mode.
Send: neph mode
Receive: neph mode sample
neph prev bkg
This command reports the previous nephelometer background values
stored on the nephelometer interface board. The following example reports
the previous nephelometer auto-ranging background values.
Send: neph prev bkg
Receive: neph prev bkg 0.00272 0.00272 0.01094 0.04357 0.17422
0.65535
set reset tape counter
This command resets the tape count to zero.
Send: set reset tape counter
Receive: set reset tape counter ok
sample rh
This command reports the sample RH, in percent. The following example
reports that the sample RH is 20%.
Send: sample rh
Receive: sample rh 20.0 %