Instruction manual
C-Link Protocol Commands
Diagnostics
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-25
Send: diag volt mb
Receive: diag volt mb 24.10 14.90 4.90 3.20 -3.20
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, and positive 3.3. Each voltage value is
separated by a space.
Send: diag volt mib
Receive: diag volt mib 24.98 14.80 -14.90 4.96 3.20
diag volt neph
This command reports the current voltage readings of the nephelometer
interface board. The following example reports the nephelometer interface
board values.
Send: diag volt neph
Receive: diag volt mib 3.3 3.3 5.0 7.2 -5.0 -7
flow
This command reports the current sample flow. The following example
reports that the sample flow is 16.69 LPM.
Send: sample flow
Receive: sample flow 16.69 LPM
flow pres
This command reports the current flow pressure for the detector, in
mmHg. The following example reports that the flow pressure is 24.1
mmHg.
Send: flow pres
Receive: flow pres 24.1 mmHg
flow temp
This command reports the current flow temperature for the detector, in
degrees C. The following example reports that the flow temperature is 22.3
°C.
Send: flow temp
Receive: flow temp 22.3 deg
io bd
This command reports the presence of the I/O expansion board (yes/no).
The following example reports the I/O expansion board is present (yes).