Instruction manual

C-Link Protocol Commands
Alarms
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-21
Send: alarm nb status
Receive: alarm nb status ok
alarm ncal status
This command reports the status of the last nephelometer auto-zero
calibration. The following example reports that the ncal is functional.
Send: alarm ncal status
Receive: alarm ncal status ok
alarm neph led cur min
alarm neph led cur max
These commands report the current nephelometer LED alarm minimum
and maximum value settings. The following example reports that the
nephelometer LED alarm minimum value is 50 mA.
Send: alarm neph led cur min
Receive: alarm neph led cur min 50.0 mA
set alarm neph led cur min value
set alarm neph led cur max value
These commands set the nephelometer LED alarm minimum and
maximum values to value, where value is a floating-point number
representing nephelometer LED alarm limits in mA. The following
example sets the nephelometer LED alarm maximum value to 75.0 mA.
Send: set alarm neph led cur min 75
Receive: set alarm neph led cur min 75 ok
alarm neph rh min
alarm neph rh max
These commands report the current nephelometer relative humidity alarm
minimum and maximum value settings. The following example reports
that the nephelometer relative humidity alarm minimum value is 5.0%.
Send: alarm neph rh min
Receive: alarm neph rh min 5.0 %
set alarm neph rh min value
set alarm neph rh max value
These commands set the nephelometer relative humidity alarm minimum
and maximum values to value, where value is a floating-point number
representing nephelometer relative humidity alarm limits in percent. The
following example sets the nephelometer relative humidity alarm maximum
value to 95%.
Send: set alarm neph rh min 95
Receive: set alarm neph rh min 95 ok