Instruction manual

C-Link Protocol Commands
Alarms
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-15
alarm amb temp min
alarm amb temp max
These commands report the current ambient temperature alarm minimum
and maximum value settings. The following example reports that the
ambient temperature alarm minimum value is 4.0 °C.
Send: alarm amb temp min
Receive: alarm amb temp min 4.0 degC
set alarm amb temp min value
set alarm amb temp max value
These commands set the ambient temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing ambient temperature alarm limits in °C. The following
example sets the ambient temperature alarm maximum value to 38 °C.
Send: set alarm amb temp max 38
Receive: set alarm amb temp max 38 ok
alarm baro pres min
alarm baro pres max
These commands report the current barometric pressure alarm minimum
and maximum value settings. The following example reports that the
barometric pressure alarm minimum value is 500.0 mmHg.
Send: alarm baro pres min
Receive: alarm baro pres min 500.0 mmHg
set alarm baro pres min value
set alarm baro pres max value
These commands set the barometric pressure alarm minimum and
maximum values to value, where value is a floating-point number
representing barometric pressure alarm limits in millimeters of mercury.
The following example sets the barometric pressure alarm maximum value
to 800.0 mmHg.
Send: set alarm baro pres max 800
Receive: set alarm baro pres max 800 ok
alarm bench status
This command reports the current status of the bench as FAIL or OK. The
following example reports that the bench status is OK.
Send: alarm bench status
Receive: alarm bench status ok
alarm beta count min
alarm beta count max