Instruction manual

C-Link Protocol Commands
Measurements
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual B-11
Command Description Page
tape counter Reports the number of counts corresponding to the filter tape
movement
B-27
tape to ncal ratio Reports/sets the ratio of filter tape changes per
nephelometer auto-zero
B-43
temp comp Reports/sets temperature compensation to actual or standard B-50
temp std Reports/sets the standard temperature B-50
temp th Reports/sets the temperature threshold for the detector B-51
thermal mass coef Reports/sets the current thermal mass coefficient B-40
time Reports/sets current time (24-hour format) B-53
tz Reports/sets the timezone string for the NTP server B-60
up Simulates pressing up pushbutton B-44
vac mass coef Reports/sets the current vacuum mass coefficient B-41
vac pres Reports the current vacuum pressure B-27
vac pres span Reports/sets the vacuum pressure span B-43
vac pres offset Reports the vacuum pressure offset B-44
vf pres cal def Sets vacuum and flow pressure calibration defaults B-44
avg24 time
This command reports the 24-hour averaging time in hours and minutes.
The following example shows that the averaging time is set to 18:30.
Send: avg24 time
Receive: avg24 time 18:30
set avg24 time hh:mm
hh = hours (01 to 23)
mm = minutes (01 to 59)
This command sets the 24-hour averaging time to start averaging the
concentration from hh:mm until next 24 hours. It is then set to next 24
hours automatically. The following example sets the 24-hour averaging
time to 17:50.
Send: set avg24 time 17:50
Receive: set avg24 time 17:50 ok
Measurements