Instruction manual
Model 5014i to 5030i Upgrade Procedures
SHARP Optics Assembly Installation
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-9
Use the following procedure for SHARP external valve installation.
1. Remove the SHARP Optics Assembly port protection caps and plugs.
2. Align the radius tube adapter with the open port on the bottom of the
optics assembly as per Figure 10–9 and gently push down and then
screw down the two tabs using the captive hardware.
Figure 10–9. Optics Installation
3. Connect the cable assembly from the rear panel of the Optical assembly
and secure it to the optics connector that was installed in Step 3 of this
procedure and as shown in Figure 10–10.
Figure 10–10. Connecting the Optics to the Optics Connector
SHARP Optics
Assembly
Installation