Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 10-1
Chapter 10
Model 5014i to 5030i Upgrade
Procedures
This chapter describes the 5014i to 5030i upgrade procedures. For details,
see the following topics:
● “Instrument Preparation” on page 10-1
● “SHARP Connection Bracket Assembly Installation” on page 10-2
● “Radius Tube Adapter Installation” on page 10-5
● “SHARP External Valve Installation” on page 10-7
● “SHARP Optics Assembly Installation” on page 10-9
● “Upgrade of the Heater Tube Assembly” on page 10-10
● “Model Label Change” on page 10-11
● “Power On Instrument” on page 10-11
Use the following procedure to prepare the instrument.
1. If upgrading at an installed site, take the appropriate measures to secure
the sample downtube and remove the instrument heater assembly and
set aside.
2. From Instrument Panel, select Main Menu, select Instrument Controls
and then Service Mode. Toggle to “Service Mode On”.
3. From Instrument Panel, select Filter Tape, and then Manual, and select
the BENCH to open.
4. From Instrument Panel, select Diagnostics, Instrument Configuration,
and change the DETECTOR configuration from BETA to SHARP.
5. Turn instrument power off with the measurement head in the open
position.
Instrument
Preparation