Instruction manual

Servicing
Delrin Nut Replacement
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-49
Figure 7–28. Replacing the SHARP Optics Board
Use the following procedure to replace the Delrin nut (Figure 7–29).
Equipment Required:
Delrin nut
Philips screwdriver
Set of standard hex drives
Adjustable wrench
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the Pump Flow OFF: From the Main Menu, choose Instrument
Controls > Set Flow/Pump > Pump. Press to toggle and set the
pump flow to OFF.
Delrin Nut
Replacement
Loosen screws
Loosen screws