Instruction manual

Servicing
SHARP RH/Temp Sensor Replacement
7-46 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Use the following procedure to replace the SHARP RH/Temp sensor
(Figure 7–26).
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF, unplug the power cord, and remove the
nephelometer from the base instrument.
2. Remove the two side cover screws, and remove cover.
3. Disconnect the RH/Temp connector from the SHARP MIB board
“TEMP/RH”.
4. Remove the two socket-head screws from the RH/Temp board
assembly from the upper optic chamber endplate.
5. Remove the RH/Temp gasket from the old RH/Temp board assembly
and add it to the new RH/Temp board assembly.
6. Slide in the new RH/Temp assembly by following the previous steps.
7. Calibrate the RH/Temp sensor. Refer to the “Calibration” chapter in
this manual.
SHARP RH/Temp
Sensor
Replacement