Instruction manual

Servicing
Radius Tube Replacement
7-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Disconnect cables from the measurement interface board and
motherboard.
3. Unscrew the four mounting screws and lift out the transformer
assembly.
4. Install the new transformer by following the previous steps in reverse.
Figure 7–20. Replacing the Transformer
Use the following procedure to replace the radius tube (Figure 7–21).
Equipment Required:
Radius tube assembly
Hex driver, 9/16-inch
Hex driver, 7/64-inch
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
Radius Tube
Replacement
Unfasten Screws (4)
Unhook Long Cable from
Transformer to back of
Measurement Interface Board
Unhook Short Cable from
Transformer from Common
Electronic Board “AC INT BD”