Instruction manual
Servicing
Flow RTD Replacement
7-36 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Figure 7–18. Replacing the Detector Assembly
Use the following procedure to replace the flow resistive temperature
diction (RTD) (see Figure 7–19).
Equipment Required:
Sample flow RTD
Adjustable wrench
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1. Turn the instrument OFF, unplug the power cord, and remove the
cover.
2. Refer to “Removing the Measurement Case Assembly and Lowering
the Partition Panel” in this chapter to lower the partition panel, then
proceed to the next step below.
3. Loosen the kynar fittings using the wrench and slide out sample flow
RTD.
Flow RTD
Replacement
Unfasten Screws (2)
Unhook SHV Connector