Detector Amplifier Assembly Replacement
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 7-15
5. Remove the four fan mounting screws and remove the fan.
6. Install a new fan following the previous steps in reverse.
Figure 7–6. Replacing the Fan
Use the following procedure to replace the detector/preamplifier assembly
(Figure 7–7).
Equipment Required:
Detector amplifier assembly
Philips screwdriver
Adjustable wrench
WARNING The detector pre-amp assembly has a power backup built in.
“DO NOT OPEN” caution handeling. Backup power will dissipate after 8
hours.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Turn the instrument OFF and unplug the power cord.
Detector Amplifier
Assembly
Replacement
Mounting Screws (4)
Filter
Fan Guard
Keps Nuts (4)