Instruction manual
Preventive Maintenance
Cleaning the Inlets
5-4 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
Figure 5–2. European PM
10
Inlet
PM
10
size selective inlet (1 m³/h) for 5030i Beta:
● Same construction as the EN reference inlet, but with 1 m³/h
● With special surface protection using an aluminium EMATAL surface
● With special water separator and easy exchangeable impactor baffle
plate
● Standard tube connection: d = 16 mm for connecting a continuous
monitor
● Easy converting to PM
2.5
using the PM
2.5
nozzle plate (8 nozzles)
● Easy converting to PM
1
using the PM
1
nozzle plate (8 nozzles)
If a Sharp-Cut Cyclone (SCC) or a BGI Very Sharp-Cut Cyclone is being
used for PM
2.5
sampling (Figure 5–3), once every 2-3 weeks the dust pot
should be unscrewed, emptied, and cleaned with a lint-free cloth. Once
every three months, the entire SCC assembly should be inspected and
cleaned, O-rings checked for distortion, cracks, fraying, or other problems.
Replace O-rings as necessary.
Cyclone
Maintenance
Nozzle (8)
Baffle Plate
Safety Snap-Locks
Tube (d = 16/14)