Instruction manual
Calibration
Calibration Procedure
4-14 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
● In the Main Menu, choose Service > Neph Calibration > Neph Temp.
CALIBRATE NEPH TEMP:
NEPH TEMP: 0.0
o
C
OFFSET: 0.0
o
C
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Source Level screen allows the user to view and
calibrate the source level. This should be within 60-70 mA. If it is outside
of this range, please adjust the % SRC Level until the IRED reads 65 mA
● In the Main Menu, choose Service > Neph Calibration > Neph Src
Level.
SOURCE LEVEL SET:
IRED: 65 mA
SRC LEVEL: 32 %
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Nephelometer Source
Level