Instruction manual

Calibration
Calibration Procedure
4-10 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
3. Cut/break the filter tape and remove from the bench. Then insert the
Foil Holder from the calibration kit and press while maintaining
a slight sideward pressure on the filter holder.
MASS COEF: 7000.0
FOIL VALUE: 1328 3μg
REMOVE TAPE, INSERT FOIL
HOLDER, TO CLOSE BENCH
RANGE AVG DIAGS ALARM
4. Insert the Null/Zero foil with the label side up and press to
proceed with the foil calibration, or press to stop the foil.
The auto zero procedure of mass will begin and last approximately 270
seconds. When the zero is complete, the beta attenuation measurement
will remain closed and an average Beta count rate is displayed.
MASS COEF: 7000.0
FOIL VALUE: 1328 3μg
Beta Avg 12022 1/sec
REMOVE NULL FOIL
TO CONTINUE
RANGE AVG DIAGS ALARM
5. Remove the zero/null foil and insert with the Span foil (e.g., 1328 ug)
and press to begin the span calibration, or press to stop
the foil.
MASS COEF: 7000.0
FOIL VALUE: 1328 3µg
INSERT NULL FOIL
TO START ZERO CAL
RANGE AVG DIAGS ALARM
MASS COEF: 7000.0
FOIL VALUE: 1328 3µg
Beta Avg 12016 1/sec
PLEASE WAIT 265 sec
RANGE AVG DIAGS ALARM