Instruction manual

Operation
Service Menu
3-102 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
The Display Pixel Test screen is used to test the LCD display, by toggling
between all pixels on and all pixels off to ensure that they are functioning
properly. The display pixel test screen is visible only when the instrument is
in service mode. For more information on the service mode, see “Service
Mode” earlier in this chapter.
In the Main Menu, choose Service > Display Pixel Test.
DISPLAY PIXEL TEST:
DURING TEST PRESS OR ¨x
TO EXIT, TO TOGGLEXXXXX
BEGIN TEST
GO BACK TO MENU
RANGE AVG DIAGS ALARM
The Restore User Defaults screen is used to reset the user calibration and
configuration values to factory defaults. The restore user defaults screen is
visible only when the instrument is in service mode. For more information
on the service mode, see “Service Mode” earlier in this chapter.
In the Main Menu, choose Service > Restore User Defaults.
The Restore User Defaults screen is used to reset the user calibration and
configuration values to factory defaults. The restore user defaults screen is
visible only when the instrument is in service mode. For more information
on the service mode, see “Service Mode” earlier in this chapter.
In the Main Menu, choose Service > Restore User Defaults.
Display Pixel Test
Restore User Defaults
Restore Factory
Defaults
RESTORE USER DEFAULTS:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE USER DEFAULTS:
RESTORE
ARE YOU SURE YOU WANT TO?x
PRESS ¨ TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM