Instruction manual
Operation
Service Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-101
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5 ª
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
● In the Main Menu, choose Service > Analog Input Cal > select
Channel > Calibrate Zero.
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.04 V
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
● In the Main Menu, choose Service > Analog Input Cal > select
Channel > Calibrate Full Scale. (Hook up a voltage source of 10 V
to the analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 9.84 V
SET TO: 10.00 V
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
Analog Input Calibrate
Zero
Analog Input Calibrate
Full-Scale