Instruction manual

Operation
Service Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-101
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5 ª
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
In the Main Menu, choose Service > Analog Input Cal > select
Channel > Calibrate Zero.
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 0.04 V
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM
The Analog Input Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
In the Main Menu, choose Service > Analog Input Cal > select
Channel > Calibrate Full Scale. (Hook up a voltage source of 10 V
to the analog input channel.)
ANALOG INPUT CAL: SPAN
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 9.84 V
SET TO: 10.00 V
CALIBRATE TO VALUE
RANGE AVG DIAGS ALARM
Analog Input Calibrate
Zero
Analog Input Calibrate
Full-Scale