Service Menu
3-98 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
● In the Main Menu, choose Service > Nephelometer Calibration.
NEPHELOMETER CALIBRATION:
>NEPH RH 4.0 %
NEPH TEMP 2.6
o
C
NEPH SRC LEVEL 35 %
RANGE AVG DIAGS ALARM
The Nephelometer RH screen allows the user to view and calibrate the
nephelometer relative humidity.
● In the Main Menu, choose Service > Neph Calibration > Neph RH.
CALIBRATE NEPH RH:
NEPH RH: 50.0 %
OFFSET: 3.5 %
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Temperature screen allows the user to view and
calibrate the nephelometer temperature sensor calibration.
● In the Main Menu, choose Service > Neph Calibration > Neph
Temp.
CALIBRATE NEPH TEMP:
NEPH TEMP: 25.0
o
C
OFFSET: 2.2
o
C
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Nephelometer Source Level screen allows the user to view and
calibrate the source level.
● In the Main Menu, choose Service > Neph Calibration > Neph Src
Level.
Nephelometer RH
Nephelometer
Temperature
Nephelometer Source
Level