Instruction manual

Operation
Service Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-95
Note This value is set in the factory and should not be adjusted.
In the Main Menu, choose Service > Mass Calibration > Alpha Eff
Coefficient.
ALPHA EFFICIENCY COEF:
CURRENTLY: 0.120000
SET TO: 0.120001 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Detector Calibration submenu is used to view and set the detector
calibration. The detector calibration screen is visible only when the
instrument is in service mode. For more information on the service mode,
see “Service Mode” earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician on a defined service interval and at least once per year, or if the
detector is replaced.
In the Main Menu, choose Service > Detector Calibration.
DETECTOR CALIBRATION:
>AUTO
MANUAL
RANGE AVG DIAGS ALARM
The Auto Detector Calibration screen allows the user to conduct a detector
optimization routine that will establish the optimal plateau for detector
performance. This process takes approximately 30 minutes and will set a
high voltage and thresholds for alpha and beta reference counts.
In the Main Menu, choose Service > Detector Calibration > Auto.
Detector Calibration
Auto Detector Calibration