3-94 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
MASS COEFFICIENT:
>AUTO
MANUAL
RANGE AVG DIAGS ALARM
The Auto Mass Coefficient screen allows the user to conduct a mass foil
calibration.
Note After pressing a series of screens will walk you through the
mass foil calibration procedure. For details, see “Auto Mass Coefficient” on
page 4-9. ▲
● In the Main Menu, choose Service > Mass Calibration > Mass >
Auto.
MASS COEF: 7000.0
FOIL VALUE: 1442 3μg
TO SAVE FOIL VALUE
AND OPEN BENCH
RANGE AVG DIAGS ALARM
The Manual Mass Coefficient screen allows the user to adjust the mass
coefficient used in beta attenuation.
● In the Main Menu, choose Service > Mass Calibration > Mass >
Manual.
MASS COEFFICIENT:
CURRENTLY: 7000
SET TO: 6992 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Alpha Efficiency Coefficient screen allows the user to view and set the
alpha efficiency value.
Auto Mass Coefficient
Manual Mass Coefficient
Alpha Efficiency
Coefficient