Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-93
THERMAL MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0023.000 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Vacuum Mass coefficient screen allows the user to view and set the
vacuum coefficient using manual calibration mode. (Factory set.)
● In the Main Menu, choose Service > Mass Calibration > Vacuum
Coefficient.
VACUUM MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0002.000 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Barometer Mass Coefficient screen allows the user to view and set the
barometer coefficient using manual calibration mode. (Factory set.)
● In the Main Menu, choose Service > Mass Calibration > Baro
Coefficient.
BARO. MASS COEFFICIENT:
CURRENTLY: 0.000000
SET TO: 0.000540 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Mass Coefficient submenu allows the user to view and set the mass
coefficient using manual or auto mode. (Factory set.)
● In the Main Menu, choose Service > Mass Calibration > Mass
Coefficient.
Vacuum Mass Coefficient
Barometer Mass
Coefficient
Mass Coefficient