Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-87
CALIBRATE AMBIENT RH:
REL HUMIDITY: 40.0 %
OFFSET: 0.0 %
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure/Vacuum Calibration submenu allows the user to view and
calibrate the pressure sensor and vacuum flow sensor. The pressure/vacuum
calibration submenu is visible only when the instrument is in service mode.
For more information on the service mode, see “Service Mode” earlier in
this chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Pres/Vacuum Calibration.
PRES/VACUUM CALIBRATION:
>BARO PRES
VAC/FLOW
RANGE AVG DIAGS ALARM
The Barometer Pressure Calibration submenu is used to calibrate the
barometer pressure offset, span or reset default values.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres.
BARO PRES CALIBRATION:
>OFFSET 0.0
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
Pressure/Vacuum
Calibration
Barometer Pressure
Calibration