Instruction manual

Operation
Alarms Menu
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-83
minimum and maximum instant PM concentration screens function the
same way.
In the Main Menu, choose Alarms > Conc Alarms > Inst PM > Min
or Max.
INSTANT SHARP CONC:
ACT MIN: 0.0
SET MIN: xxxxx-10 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Instant SHARP screen displays the current instant SHARP
concentration and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from -10 to 10000 μg/m
3
or -.01 to 10
mg/m
3
. If the instant SHARP concentration goes beyond either the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Conc Alarms > Inst SHARP.
INSTANT SHARP CONC:
ACTUAL 0.0
>MIN -10.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Instant SHARP Concentration alarm limit screen is used to
change the minimum instant SHARP concentration alarm limit. The
minimum and maximum instant SHARP concentration screens function
the same way.
In the Main Menu, choose Alarms > Conc Alarms > Inst SHARP >
Min or Max.
Instant SHARP
Min and Max Instant
SHARP Concentration
Limits