Instruction manual
Operation
Alarms Menu
3-82 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
AVERAGE NEPH CONC:
ACTUAL 0.0
>MIN 0.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Average Nephelometer Concentration alarm limit screen is
used to change the minimum average nephelometer concentration alarm
limit. The minimum and maximum average nephelometer concentration
screens function the same way.
● In the Main Menu, choose Alarms > Conc Alarms > Avg Neph >
Min or Max.
AVERAGE NEPH CONC:
ACT MIN: 0.0
SET MIN: xxxxx100 ?
§¨ MOVE CURSOR
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Instant PM screen displays the current instant PM concentration and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from -10 to 10000 μg/m
3
or -.01 to 10 mg/m
3
. If the instant PM
concentration goes beyond either the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on the
Run screen and in the Main Menu.
● In the Main Menu, choose Alarms > Conc Alarms > Inst PM.
INSTANT PM CONC:
ACTUAL 0.0
>MIN -10.0
MAX 10000
RANGE AVG DIAGS ALARM
The Minimum Instant PM Concentration alarm limit screen is used to
change the minimum instant PM concentration alarm limit. The
Min and Max Average
Nephelometer
Concentration Limits
Instant PM
Min and Max Instant PM
Concentration Limits