Instruction manual
Operation
Alarms Menu
3-72 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
The RH/Temperature Alarms submenu allows the user to view and set the
relative humidity or temperature alarm parameters.
โ In the Main Menu, choose Alarms > RH/Temperature Alarms.
RH/TEMPERATURE ALARMS:
>AMBIENT RH OK
SAMPLE RH OK
AMBIENT TEMP OK
FLOW TEMP OK
BOARD TEMP OKยช
RANGE AVG DIAGS ALARM
The Ambient RH screen displays the current ambient relative humidity
value and sets the minimum and maximum alarm limits. Acceptable alarm
limits range from 5 to 95%. If the ambient RH reading goes beyond the
minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
โ In the Main Menu, choose Alarms > RH/Termperature Alarms >
Ambient RH.
AMBIENT RH:
ACTUAL 27.4 %
>MIN 5.0 %
MAX 95.0 %
RANGE AVG DIAGS ALARM
The Minimum Ambient RH alarm limit screen is used to change the
minimum ambient RH alarm limit. The minimum and maximum ambient
RH screens function the same way.
โ In the Main Menu, choose Alarms > RH/Temperature Alarms >
Ambient RH > Min or Max.
RH/Temperature
Alarms
Ambient RH
Min and Max Ambient RH
Limits