Instruction manual
Operation
Volumetric Conditions
3-50 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
analog output channel. See Table 3–4 for a list of choices. The minimum
and maximum output value screens function the same way. The following
example shows the set minimum value screen.
● In the Main Menu, choose Instrument Controls > IO Configuration
> Analog Output Config > select Channel > Set Minimum or
Maximum Value.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.5 % ?
©ª CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Table 3–4. Analog Output Zero to Full-Scale
Output Zero (0)% Value Full-Scale 100% Value
PM Zero (0) Range Setting
AVG PM Zero (0) Range Setting
Neph Zero (0) Range Setting
SHARP Zero (0) Range Setting
AVG SHARP Zero (0) Range Setting
AVG Neph Zero (0) Range Setting
Barometer Pressure User-set alarm min value User-set alarm max value
Vacuum User-set alarm min value User-set alarm max value
Flow Pressure User-set alarm min value User-set alarm max value
Ambient RH User-set alarm min value User-set alarm max value
Sample RH User-set alarm min value User-set alarm max value
Ambient Temp User-set alarm min value User-set alarm max value
Board Temp User-set alarm min value User-set alarm max value
Flow Temp User-set alarm min value User-set alarm max value
Flow Vol User-set alarm min value User-set alarm max value
Beta User-set alarm min value User-set alarm max value
Beta Raw User-set alarm min value User-set alarm max value
Alpha Raw User-set alarm min value User-set alarm max value
Mass -200 5,000
Beta Zero 0 65,000