Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-47
Not all of the I/O available in the instrument are brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional “Terminal Block and Cable Kits”.)
In the Main Menu, choose Instrument Controls > I/O Configuration
> Digital Input Settings.
DIGITAL INPUT SETTINGS:
>1 NOP AOUTS TO FS
2 NOP AOUTS TO ZERO
3 NOP PUMP CNTRL
4 NOP PRES CNTRL
5 NOP NONE ª
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital input pin and ground is
normally open and closes to trigger the digital input action.
Press to toggle and set the logic state to normally open or
normally closed.
DIGITAL INPUT SETUP:
>LOGIC STATE OPEN
INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
The Instrument Action submenu allows the user to choose the instrument
action that is assigned to the selected digital input.
In the Main Menu, choose Instrument Controls > I/O Configuration
> Digital Input Settings > select Relay > Instrument Action.
Logic State
Instrument Action