Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-47
Not all of the I/O available in the instrument are brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional “Terminal Block and Cable Kits”.) ▲
● In the Main Menu, choose Instrument Controls > I/O Configuration
> Digital Input Settings.
DIGITAL INPUT SETTINGS:
>1 NOP AOUTS TO FS
2 NOP AOUTS TO ZERO
3 NOP PUMP CNTRL
4 NOP PRES CNTRL
5 NOP NONE ª
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital input pin and ground is
normally open and closes to trigger the digital input action.
● Press to toggle and set the logic state to normally open or
normally closed.
DIGITAL INPUT SETUP:
>LOGIC STATE OPEN
INSTRUMENT ACTION
RANGE AVG DIAGS ALARM
The Instrument Action submenu allows the user to choose the instrument
action that is assigned to the selected digital input.
● In the Main Menu, choose Instrument Controls > I/O Configuration
> Digital Input Settings > select Relay > Instrument Action.
Logic State
Instrument Action