Instruction manual
Operation
Volumetric Conditions
Thermo Fisher Scientific Model 5030i SHARP Instruction Manual 3-45
Note The digital outputs may take up to one second after the assigned
state occurs to show up on the outputs. ▲
● In the Main Menu, choose Instrument Controls > I/O Configuration
> Output Relay Settings.
OUTPUT RELAY SETTINGS:
>1 NOP CONC ALARM
2 NOP LOCAL/REMOTE
3 NOP UNITS
4 NOP GEN ALARM
5 NOP NONE ª
RANGE AVG DIAGS ALARM
The Logic State menu item is used to change the selected I/O relay to
either normally open or normally closed. The default state is open, which
indicates that a relay connected between the digital output pin and ground
is normally open and closes to trigger the digital output action.
● Press to toggle and set the logic state to normally open or
normally closed.
OUTPUT RELAY SETUP:
>LOGIC STATE OPEN
INSTRUMENT STATE
RANGE AVG DIAGS ALARM
The Instrument State submenu allows the user to select the instrument
state that is assigned to the selected relay output. A submenu lists signal
types of either alarms or non-alarm to choose from.
● In the Main Menu, choose Instrument Controls > I/O Configuration
> Output Relay Settings > select Relay > Instrument State.
CHOOSE SIGNAL TYPE:
>ALARMS
NON-ALARM
RANGE AVG DIAGS ALARM
Logic State
Instrument State