Instruction manual

Operation
Volumetric Conditions
3-38 Model 5030i SHARP Instruction Manual Thermo Fisher Scientific
INSTRUMENT ID:
CURRENTLY: 14
SET TO: 16 ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Gesytec Serial Number screen is used to set the serial number for the
Gesytec protocol. The Gesytec serial number defaults to zero. For more
information about the serial number, see Appendix D, “Gesytec (BH)
Protocol”.
In the Main Menu, choose Instrument Controls > Communication
Settings > Gesytec Serial No.
GESYTEC SERIAL NO:
CURRENTLY: 0
SET TO: 1 ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Communication Protocol screen is used to change the protocol for
serial communications. Possible choices include: C-Link, MODBUS,
Gesytec, and Streaming Data.
In the Main Menu, choose Instrument Controls > Communication
Settings > Communication Protocol.
COMMUNICATION PROTOCOL:
CURRENTLY: CLINK
SET TO: STREAMING ?
©ª CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Streaming Data Configuration menu is used to allow for configuration
of the 8 streaming data output items, streaming interval, current data
format, and current timestamp setting. The Choose Stream Data submenu
displays a list of the analog output signal group choices to choose from.
Gesytec Serial Nunber
Communication Protocol
Streaming Data Configuration