Instruction manual

STP-200/300/400 Series Instruction Manual
4.1.2 Rear Panel
(See Figure 4.2.)
(11) AC POWER Connector (CON30)
For primary power input.
(12) MAIN POWER Breaker
Switches ON/OFF the primary power.
A metal fitting is attached to secure the breaker at the OFF
position.
(13) P. CONNECTOR (CON20)
For connection of the STP connection cable.
(14) REMOTE CONNECTOR (CON40)
For remote control.
For details of the REMOTE CONNECTOR (14), see Section 8, “Remote
Input/Output Signal Connector.”
(15) "MANUAL/REMOTE" Changeover Switch
When setting to MANUAL
, only start, stop or other operations can
be performed with the switches on the STP control unit front
panel.
When setting to REMOTE, only start, stop or other operations can
be performed by inputting the remote signal.
(16) EXT. BATTERY Connector (CON90)
For connection of an external battery.
When using an external battery, always disconnect the internal
battery connector (See Section 10, "External Battery" for details).
(17) Ground Terminal
For connection of the ground cable between the STP pump and the
STP control unit.
(18) Air Cooling Fan
For cooling the inside of the STP control unit.
4-3