4 Measurement
52
Relay operation
Check the operations of contact outputs R1 and R2.
Checking the contact output R1
1.
Press the UP/DOWN key to display “rLY” on the sub-display.
2. Press the ENT key.
“rLY1” is shown on the measurement display and “rLY” on the sub-display.
3. Press the ENT key.
The relay lamp R1 lights up and the relay is turned on, normally.
Press the ENT key to switch the relay ON/OFF to check the operation.
4. Press the MEAS key.
The instrument returns to the user check menu entrance.
Checking the contact output R2
You can check R2 in the same way as R1. But, press the UP/DOWN key to display “rLY2” on
the sub-display after the step 2.
Measured value and temperature
1.
Press the UP/DOWN key in the user check menu to display “diSP” on the sub-
display.
2. Press the ENT key.
The current measured value is shown on the measurement display and the current
temperature on the sub-display.
3. Press the MEAS key.
The instrument returns to the user check menu entrance.
4.6.2 Instrument reset
This operation allows you to initialize all the settings and restart the instrument.
Then perform the steps in “ Initial set up” (page 16) in the automatically opened basic setup
menu.
Calibration values are also reset. Be sure to perform the steps in “4.3 Calibration” (page 43)
before use.
Operation
1.
Enter the user check menu referring to “How to open the user check menu” (page
51).
”USr” is shown on the measurement display and “LEd” on the sub-display.
2. Press the DOWN key to display “init” on the sub-display, and then press the ENT
key.
The sub-display shows “no” and starts blinking.
3. Press the DOWN key to select “YES”, and then press the ENT key.
The instrument is reset to the state before factory shipment.
The sub-display shows “good” with flashing, and the instrument restarts.
Tip
z The initial settings are shown in “Table 1 Basic setup items” (page 16).
z To return to the user check menu entrance without resetting, select “no” in step 3. and press the
ENT key.