Instruction manual
4 Measurement
51
4.6 User check menu
The user check menu allow you to check the instrument status, and to reset it to the default.
Operate it from the user check menu.
How to open the user check menu
Start the operation it from the measurement mode.
1. Hold down the MNT key until the MNT lamp lights up.
The instrument enters the maintenance mode, and “SEt” is shown on the measurement
display.
2. Press the DOWN key to display “USr” on the measurement display, and then press
the ENT key.
The instrument opens the user check menu, and “LEd” is shown on the sub-display.
Refer to the below instructions for perform each of the available operations in accordance
with the following description.
4.6.1 Status check
You can check the four items shown below.
z LEDs
z Transmission output
z Relay operation
z Measured value and temperature
Operation
Perform the following operations after opening the user check menu.
LEDs
1.
Press the UP/DOWN key, if necessary, to display “LEd” on the sub-display, and
press the ENT key.
All the LED will light up normally.
2. Check that LEDs light up without problems.
3. Press the MEAS key.
The instrument returns to the user check menu entrance.
Transmission output
You can output an electric current of 4 mA, 12 mA and 20 mA from the transmission output.
1. Press the UP/DOWN key to display “Cur” on the sub-display.
2. Press the ENT key.
“4-20” is shown on the measurement display and “4” on the sub-display outputting a 4 mA
transmission output.
3. To switch the current output, press the UP/DOWN key to select the value
corresponding to the current (mA) to be output from 4, 12, or 20.
4. Check that the electric current is being output properly.
5. Press the MEAS key.
The instrument returns to the user check menu entrance.