4.3 Calibration
Three types of calibration are provided, namely temperature calibration, adjustment of
conductivity zero point's compensation coefficient, and adjustment of cell constant
compensation coefficient.
Table 7 Calibration setting items
Note
Be sure to perform calibration after the power is turned ON for the first time.
Precautions
z Previous value hold is used for transmission output during the maintenance mode.
z Contact outputs are in the following states during the maintenance, depending on the
selected control target.
z OFF when the control target is set to non, Cond or Err.
z ON when the control target is set to Hold.
z FAIL when the control target is set to FAiL and a system error occurs.
Display Item Description Reference Remarks
Temperature calibration
For temperature calibration;
Set the deviation from the
sample temperature only.
page 44
Adjustment of conductivity zero
point's compensation coefficient
(Adjust)
Conductivity zero point's
compensation coefficient
page 44
Only when
instrumental errors
exist.
Adjustment of cell constant
compensation coefficient (Span)
Cell constant compensation
coefficient
page 45
Only when
instrumental errors
exist.