4 Measurement
42
Hold value
Set the hold value to be used if preset hold is enable.
1. Open the basic setup menu and press the UP/DOWN keys to display the zero
setting of transmission output on the panel, referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for transmission output range's zero point setting, “mG.0”.
2. Press the DOWN key to display “PrES” on the measurement display, and then
press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to specify a Cond value to output as the hold value.
UP key : Increments the value by the minimum display unit.
DOWN key : Decrements the value by the minimum display unit.
The setting range is the measurement range plus 10% of the measurement range.
Ex.) In the case of the measurement range 0.00
µS/cm to 20.00 µS/cm
The setting range of hold value: 0.00
µS/cm to 22.00 µS/cm
4. Press the ENTER key to establish the setting.
Tip
z To set another item of transmission output setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.
Transmission output limit
Select the transmission output setting, whether values outside the range of 4 mA to 20 mA are
output or not.
1. Open the basic setup menu and press the UP/DOWN keys to display the zero
setting of transmission output on the panel, referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for transmission output range's zero point setting, “mG.0”.
2. Press the DOWN key to display “C.cut” on the measurement display, and then
press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the transmission output limit ON/OFF.
UP key: YES:Values under the zero point or over the span point are not output.
DOWN key:no: Values below the zero point or over the span point are output.
4. Press the ENTER key to establish the setting.
Tip
z To set another item of transmission output setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.