Instruction manual

4 Measurement
41
Span point of the transmission output range
Set the measured value corresponding to 20 mA is parallelized
1. Open the basic setup menu and press the UP/DOWN keys to display the zero
setting of transmission output on the panel, referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for transmission output range's zero point setting, “mG.0”.
2. Press the DOWN key to display “rnG.S”, and then press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the Cond value corresponding to the span point
output.
UP key : Increments the value by the minimum display unit.
DOWN key : Decrements the value by the minimum display unit.
The setting range is the measurement range plus 10% of the measurement range.
Ex.) In the case of the measurement range 0.00
µS/cm to 20.00 µS/cm
The setting range of transmission output range span point: 0.00
µS/cm to 22.00 µS/
cm
4. Press the ENTER key to establish the setting.
Note
If the zero point and span point are set to the same value, the actual output is fixed to 20 mA.
Tip
z To set another item of transmission output setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.
Hold output
Select the hold method when the mode is switched from measurement to calibration.
Tip
When the mode is switched to maintenance, the previous value hold is used as the hold value.
1. Open the basic setup menu and press the UP/DOWN keys to display the zero
setting of transmission output on the panel, referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for transmission output range's zero point setting, “mG.0”.
2. Press the DOWN key to display “Hold”, and then press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to select a setting.
YES : The previous value is used as the hold value (Previous value hold).
PrES : A preset value is used as the hold value (Preset hold).
no : Hold output is not used (Actual data is output continuously).
4. Press the ENTER key to establish the setting.
Tip
z To set another item of transmission output setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.