4.2.5 Transmission output
Set up the items related to transmission output.
Table 6 Settings for transmission output
Zero point of the transmission output range
Set the measured value corresponding to 4 mA is parallelized
1. Open the basic setup menu and press the UP/DOWN keys to display the zero
setting of transmission output (as below) on the panel, referring to “ How to
open basic setup menus” (page 19).
The setting screen is displayed on the measurement display, showing the display for
transmission output range's zero point setting, “mG.0”.
2. Press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to specify a conductivity value corresponding to the zero
point output.
UP key : Increments the value by the minimum display unit.
DOWN key : Decrements the value by the minimum display unit.
The setting range is the measurement range plus 10% of the measurement range.
Ex.) In the case of the measurement range 0.00
µS/cm to 20.00 µS/cm
The setting range of transmission output range zero point: 0.00
µS/cm to 22.00 µS/
cm
4. Press the ENTER key to establish the setting.
Tip
z To set another item of transmission output setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.
Display Item Description Reference Remarks
Zero point of the transmission
output range
Set the measured value
corresponding to 4 mA is
parallelized
page 40
Span point of the transmission
output range
Set the measured value
corresponding to 20 mA is
parallelized
page 41
Hold output Select the hold method page 41
Hold value Value of the optional hold page 42
Valid only when the
hold output is set to
“PrES”,
Transmission output limit
Select the transmission
output setting, whether values
outside the range of 4 mA to
20 mA are output or not
page 42