Instruction manual
4 Measurement
36
Control method setting
Select the limit operation, upper or lower
1. Open the basic setup menu and press the UP/DOWN keys to display the control
target setting for R1 on the panel, referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for control target setting, “1.SEL”.
2. Press the DOWN key to display “1.H-L” on the measurement display, and then
press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to select a setting.
UP key : H : Upper limit operation
DOWN key : L : Lower limit operation
4. Press the ENTER key to establish the setting.
Tip
z To set another item of relay output 1 setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.
Control value setting
Set the control value, used as the criterion for switching the relay output ON/OFF.
1. Open the basic setup menu and press the UP/DOWN keys to display the control
target setting for R1 on the panel, referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for control target setting, “1.SEL”.
2. Press the DOWN key to display “1.SEt” on the measurement display, and then
press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set a setting.
UP key : Increments the value by the minimum display unit.
DOWN key : Decrements the value by the minimum display unit.
The setting range is the measurement range.
4. Press the ENTER key to establish the setting.
Tip
z To set another item of relay output 1 setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.