Instruction manual
4 Measurement
35
Reverse output setting
Select whether or not to invert the control output of the control target.
1. Open the basic setup menu and press the UP/DOWN keys to display the control
target setting for R1 on the panel, referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for control target setting, “1.SEL”.
2. Press the DOWN key to display “1.rE” on the measurement display, and then press
the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to select a setting.
UP key : YES : The control output is reversed.
DOWN key : no : The control output is not reversed.
4. Press the ENTER key to establish the setting.
Tip
z To set another item of relay output 1 setting next, press the UP/DOWN key to display it.
z To select another setting item, press the MEAS key.
How to output FAIL during power OFF.
To output FAIL while the power supply is turned off, make the following three setting.
z Connect a cable between C and NC.
z Set the control target (1.SEL) to FAiL.
z Set the reverse output (1.rE) YES.
Note
When this function is enable, the target of the delay time setting is also reversed. To delay the FAIL
output, use the delay time setting for relay OFF.
Tip
Set the reverse output (1.rE) to no whenever the control target (I.SEL) is set to an item other than FAiL.