Instruction manual

4 Measurement
33
Control target selection
Select the control target of the contact output.
1. Open the basic setup menu and press the UP/DOWN keys to display the control
target setting for R1 (as below) on the panel, referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for control target setting, “1.SEL”.
2. Press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to select a setting.
non : No target
Cond : Measured value
Hold : During maintenance
Err : Error
FAiL : FAIL
USP : USP conductivity assessment
4. Press the ENTER key to establish the setting.
Tip
z To set another item of relay output 1 setting next, press the DOWN key to display it.
z To select another setting item, press the MEAS key.
Safety factor setting for USP criterial value
The safety factor (%) setting is provided to allow you to set a tighter criterion than USP criterial
value.
1. Open the relay output R1 setting screen referring to “How to open basic setup
menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for control target setting, “1.SEL”.
2. Display “1.USP” on the measurement display using the DOWN key, and press the
ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the safety factor.
UP key : Increments the value by 1%.
DOWN key : Decrements the value by 1%.
The setting range is 30% to 100%.
4. Press the ENT key to establish the setting.
Tip
z To set another item in the relay output R1 setting next, press the UP/DOWN key to display it.
z To return to the basic setup menu, press the MEAS key.
Current setting