Instruction manual
4 Measurement
24
4.2.2 Measurement condition
Unit selection
Select the unit for measurement values.
1. Open the measurement condition setting screen referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for unit setting, “Unit”.
2. Press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Select a unit using the UP/DOWN key.
Conv (Conventional) :
µS/cm
SI : mS/m
tdS (Total Dissolved Solids) : mg/L
4. Press the ENT key to establish the setting.
Tip
z To set another item in the measurement condition setting next, press the DOWN key to display it.
z To return to the basic setup menu, press the MEAS key.
Measurement range setting
Set the maximum value of the measurement range. the measurement range will be 0 to the
value specified here.
1. Open the measurement condition setting screen referring to “ How to open basic
setup menus” (page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for unit setting, “Unit”.
2. Display “rnG” on the measurement display using the DOWN key, and press the ENT
key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the upper limit of the range.
For the available options, refer to the table shown below:
4. Press the ENT key to establish the setting.
Tip
z To set another item in the measurement condition setting next, press the UP/DOWN key to
display it.
z To return to the basic setup menu, press the MEAS key.
← Current setting
Unit “Unit”
Cell constant “CELL”
0.01 0.1 1.0
Conv (µS/cm) 2.000 20.00 20.00 200.0 200.0 2000
SI (mS/m)
0.2000 2.000 2.000 20.00 20.00 200.0
tdS (mg/L) 2.00 20.0 20.0 200 200 2000