Instruction manual

4 Measurement
22
Cell constant compensation coefficient setting
1.
Open the sensor setting menu referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for cell constant setting, “CELL”.
2. Display “CoEF” on the measurement display using the DOWN key, and press the
ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the cell constant compensation coefficient.
UP key : Increments the value by 0.001.
DOWN key : Decrements the value by 0.001.
The setting range is 0.001 to 5.000.
4. Press the ENT key to establish the setting.
Tip
z To set another item in the sensor setting next, press the UP/DOWN key to display it.
z To return to the basic setup menu, press the MEAS key.
RTD temperature coefficient setting
1.
Open the sensor setting menu referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for cell constant setting, “CELL”.
2. Display “t_Co” on the measurement display using the DOWN key, and press the
ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Press the UP/DOWN key to set the RTD temperature coefficient.
UP key : Increments the value by 1 ppm/°C.
DOWN key : Decrements the value by 1 ppm/°C.
The setting range is 3500 to 4000 ppm/°C.
4. Press the ENT key to establish the setting.
Note
Do not change the default value (3750 ppm) when no temperature coefficient (TEMP COEF) is
specified for platinum resistance (RTD).
Tip
z To set another item in the sensor setting next, press the UP/DOWN key to display it.
z To return to the basic setup menu, press the MEAS key.