Instruction manual

4 Measurement
21
4.2.1 Sensor setting
Enter the values found on the sensor label except the cable length.
Label example:
Cell constant selection
1.
Open the sensor setting menu referring to “ How to open basic setup menus”
(page 19).
The setup item selection screen is displayed on the measurement display, showing the
display for cell constant setting, “CELL”.
2. Press the ENT key.
The sub-display starts blinking. Now you can change the setting.
3. Select the cell constant from 0.01, 0.1, and 1.0 using the UP/DOWN key.
4. Press the ENT key to establish the setting.
Note
The cell constant of the sensor is found on the lead wire at the end to be connected to the main unit.
Be sure to set this value to the main unit.
Tip
z To set another item in the sensor setting next, press the DOWN key to display it.
z To return to the basic setup menu, press the MEAS key.
MODEL TSD-01L
MFG.No. 1234567
CELL CONST 1.107
× 0.1/cm
RTD 1000 -0.26Ω
TEMP COEF 3750ppm/°C
Cell constant compensation coefficient
Cell constant, 0.1 or 0.01
Resistance compensation value for platinum
resistance (RTD), which is shown as a deviation from
1000 *
Temperature coefficient of platinum resistance (RTD) *
*:Not shown when the MODEL name is ESD
Current setting