Instruction manual

3 Preparation
18
Safety factor of USP criterial value for relay output 2 30 to 100 (%) 100
Reverse output for relay 2 YES, no no
Selection of upper or lower limit for relay 2 H, L L
Control value setting for relay 2
The setting range is 0 to the
conductivity value specified in
the measurement range setting
0.00
Selection of control width type for relay 2 d.diF, S.diF d.diF
Setting value for control width in both directions
0% to 20% of the measurement
range
4.00
Setting value for control width in one direction, width
before the control value
0% to 10% of the measurement
range
2.00
Setting value for control width in one direction, width
after the control value
0% to 10% of the measurement
range
2.00
Duration of alarm output, starting from alarm output
to the time when the contact output (between C and
NO) is turned ON (unit: seconds)
0 to 600 (seconds) 0
Duration of alarm clear, starting from alarm clear to
the time when the contact output (between C and
NO) is turned OFF (unit: seconds)
0 to 600 (seconds) 0
Transmission output setting
Cond value for zero point (4 mA) of the transmission
output range
Available options vary
depending on the settings of
cell constant “CELL” and unit
“Unit”.
0.00
Cond value for span point (20 mA) of the
transmission output range
Available options vary
depending on the settings of
cell constant “CELL” and unit
“Unit”.
20.00
Holds the previous value of transmission output
during calibration mode (except for maintenance
mode)
YES, PrES, no YES
Setting value for preset (arbitrary value) hold
Available options vary
depending on the settings of
cell constant “CELL” and unit
“Unit”.
20.00
Transmission output cut
Does not output values out of the range of 4 to 20
mA
YES, no YES
Display Description Options
Factory
setting