Instruction manual

3 Preparation
17
Reference temperature setting 5 to 95 (°C) 25
Setting of TDS conversion coefficient
When “tdS” has been set for the
unit selection: 0.30 to 1.00
0.50
Setting of the number of moving average times
(dumping factor)
1 to 20 (times) 1
Display settings
Setting of items to be displayed on the sub-display no, 1.SEt, 2.SEt, t t
Measurement range cut
Does not display values out of the conductivity
range specified in “Measurement range setting”.
YES, no YES
Decimal place for temperature values 1, 0.1, 0.01 0.1
Relay output 1 settings
Selection of the target for relay output 1 non, Cond, Hold, Err, FAiL, USP Cond
Safety factor of USP criterial value for relay output 1 30 to 100 (%) 100
Reverse output for relay 1 YES, no no
Selection of upper or lower limit for relay 1 H, L H
Control value setting for relay 1
The setting range is 0.00 to the
conductivity value specified in
the measurement range setting
20.00
Selection of control width type for relay 1
d.diF, S.diF d.diF
Setting value for control width in both directions
0% to 20% of the measurement
range
4.00
Setting value for control width in one direction, width
before the control value
0% to 10% of the measurement
range
2.00
Setting value for control width in one direction, width
after the control value
0% to 10% of the measurement
range
2.00
Duration of alarm output, starting from alarm output
to the time when the contact output (between C and
NO) is turned ON (unit: seconds)
0 to 600 (seconds) 0
Duration of alarm clear, starting from alarm clear to
the time when the contact output (between C and
NO) is turned OFF (unit: seconds)
0 to 600 (seconds) 0
Relay output 2 settings
Selection of the target for relay output 2 non, Cond, Hold, Err, FAiL, USP Cond
Display Description Options
Factory
setting