Instruction manual
3 Preparation
16
3 Preparation
When you start using the instrument for the first time after factory shipment or resetting it to
factory settings, be sure to complete the following preparation procedures.
z Wiring checking
z Initial set up
z Calibration
Wiring checking
Check the following.
z Are the power cable, sensor cable, and transmission cable properly wired?
z Are the terminal block screws firmly tightened?
z Is the fluctuation of the power supply voltage within ±10% of the rated voltage?
Initial set up
When the instrument is powered on in the factory setting state, it starts up in the Maintenance
mode.
Check the current settings referring to Table 1.
If necessary, change settings following the instructions in “4.2 Basic settings” (page 19).
To finish the initial setup, press the MEAS key repeatedly until the MEAS lamp lights up to go
to the measurement mode.
Once the initial setup is finished, the instrument starts up in the measurement mode since
then. This is the normal measurement mode.
Table 1 Basic setup items
Display Description Options
Factory
setting
Sensor settings
Selection of the cell constant of the sensor being
used.
0.01, 0.1, 1.0 (/cm) 0.01
Setting of the compensation coefficient for the cell
constant
0.001 to 5.000 1.000
Setting of the RTD temperature coefficient 3500 to 4000 (ppm/°C) 3750
Setting of the RTD resistance compensation (Ω)
±5.00 (Ω)0.00
Cable length compensation 0 to 100 (m) 10
Parameters for measurement
Unit selection
Conv (µS/cm), SI (mS/m), tdS
(mg/L)
Conv
Measurement range setting
Available options vary
depending on the settings of
cell constant “CELL” and unit
“Unit”.
20.00
Selection of calculation method of temperature
compensation
PurE, nACl, t.CoE, oFF nACl
Temperature coefficient setting
±5.00 (%/°C) 2.00