Instruction manual
2 Installation
9
2 Installation
2.1 Installation environment
To keep the instrument stable and reliable for use, install it in a place where the following
conditions are satisfied.
Instrument body
z Well-ventilated
z Ambient temperature is in the range of -5°C to 45°C
z Where the air is not hot
z Not exposed to direct sunlight
z Not exposed to direct radiant heat
z Ambient relative humidity is in the range of 20% to 85%
z The instrument is not splashed with water or chemicals
z Mechanical vibration is rare
z There is enough space for maintenance and wiring
z Dust and corrosive gases are not present
z The influence of electromagnetic fields is rare
z A place whose altitude is less than 2000 m
z The range of power supply voltage fluctuation is within 10% of the rated voltage
z A place where Overvoltage Category II is satisfied (This regulation is applied to the
electrical machinery that is powered by stationary equipment such as a switchboard.)
Sensor
z Sensors can be checked and maintained
z No bubbles appear in the measuring solution
z Measuring solution does not corrode the wetted part of the sensor.
Note
For the details, refer to the instruction manual provided with the sensor.