Instruction manual
Thermo Scientific Rotary Microtome Microm HM 340 E
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387831 - English
59
For transportation outside closed buildings,
please observe the following measures:
• Lock the handwheel.
• Turn off the main switch of the instrument.
• Remove knife or blade and store it in a safe
place.
• Remove the section waste tray.
• To lift the instrument, use the recessed grip on
the lower rear side of the instrument and the
guidance for the knife carrier.
• During transportation, do not carry the
instrument on the handwheel handles. Danger
of being injured!
• Pack the instrument into the original packing
as it offers best pre-conditions for
transportation without damage.
Caution:
Any shipping of the instrument requires original
packaging materials! Damages caused by
shipping with non-conform packaging are not
covered by the manufacturer warranty! Any
damage repairs resulting in non-conforming
package are fully charged to the sending party.
We reserve the right depending on seriousness of
damage NOT to repair.
To order original packaging materials, please
contact Thermo Fisher Scientific or your local, by
Thermo Scientific authorized, dealer.
Caution:
The user must care for a clean and safe
condition of the instrument when returning it
to an appropriate service provider.
Note:
If the original packaging is no longer available,
please contact your local Thermo Scientific
representation.