5-3 REPLACEMENT WORK
Replacing the fuses:
The two fuses of the instrument are placed on the
rear side of the instrument beside the main switch
(fig. 55.1).
• To replace the two fuses, turn off the main
switch of the instrument and unplug it.
• Then open the cover (fig. 55.2) on the right
side (see arrow) by means of a flat
screwdriver.
• Pull out the fuse holders, each of which is
marked with an arrow, and replace the fuses
with new ones.
Rating of fuses:
For power requirements 220 – 240 V:
- 2 fuses 200 mA
For power requirements 100 – 120 V:
- 2 fuses 400 mA
Fig. 55