5-3 REPLACEMENT WORK
Replacing the fuses:
The two fuses of the instrument are placed on the
rear side of the instrument beside the main switch
(fig. 55.1).
To replace the two fuses, turn off the main
switch of the instrument and unplug it.
Then open the cover (fig. 55.2) on the right
side (see arrow) by means of a flat
screwdriver.
Pull out the fuse holders, each of which is
marked with an arrow, and replace the fuses
with new ones.
Rating of fuses:
For power requirements 220 240 V:
- 2 fuses 200 mA
For power requirements 100 120 V:
- 2 fuses 400 mA
Fig. 55