4-3 POSSIBLE SOURCES OF ERRORS CAUSE AND REMOVAL
Problem
Cause
Removal
Thick-thin-sections
Blunt knife/blade.
Knife angle, therefore,
unfavorable clearance angle.
Insufficient clamping on
specimen clamping and/or
knife carrier.
Move knife and/or insert a new
one.
Try clearance angle
adjustments, until an optimal
angle can be found.
Check all screw and clamping
connections on specimen
clamping and knife carrier.
Tighten them, if necessary.
Compressions
Blunt knife/blade.
Specimen too warm.
Unfavorable clearance angle.
Cutting speed too high.
Move knife and/or insert a new
one.
Cool specimen.
Try clearance angle
adjustments until an optimal
angle can be found.
Turn handwheel slower.
"Chatter" on sections
Cutting speed too high.
Unfavorable clearance angle.
Insufficient clamping on
specimen clamping and/or
knife carrier.
Turn handwheel slower.
Try clearance angle
adjustments until an optimal
angle can be found.
Check all screw and clamping
connections on specimen
clamping and knife carrier.
Tighten them, if necessary.
No feed is carried out anymore,
and therefore, no sections are
produced
Front end position has been
reached.
Coarse feed wheel is blocked.
Move specimen backwards by
turning the coarse feed wheel.
Remove blockade on coarse
feed wheel.
Tight handwheel
- partly
- during the entire movement
Debris and section waste
between microtome and base
plate.
Dirty link block.
Remove section waste and
clean microtome.
Contact a service technician.
Instrument cannot be turned on
Power cord not correctly
connected
Defective main fuse
Check power cord
Check/replace the fuses in the
mains plug unit
Note:
In case of malfunctions and/or service work, please turn off the instrument and contact your
local dealer.